1) The MAXimum amplitude is dependent on the offset.
2) The MAX and MIN arguments should not be used in a
program message containing an OFFSet command, since
these values are evaluated during parsing, based on the
current value of the offset.
OFFSet <offset>
The offset command is used to set the DC offset of the
output waveform. Note that the amplitude and the offset
are limited by the relation:
Peak Amplitude + |Offset| ≤ 5V
Arguments
Type: Numeric
Units: V, mV
Range: 10mV to 5V
Rounding: To 10mV
Command Type: Setting or Query
Setting
Syntax: OFFSet<ws><offset>[units]
OFFSet<ws>MINimum|MAXimum
Examples: OFFS 2.5
OFFS 2.5V
OFFS MAX
Query
Syntax: OFFSet?[<ws>MINimum|MAXimum]
Examples: OFFS?
OFFS? MAX