9
Z Accuracy:
As shown in table 1.
C Accuracy:
Cxf
Zx
=
π
2
1
C
Ae
= Ae of C
f : Test Frequency (Hz)
Cx : Measured Capacitance Value (F)
|Zx| : Measured Impedance Value ()
Accuracy applies when Dx (measured D value) 0.1
When Dx > 0.1, multiply C
Ae
by
2
1 Dx+
Example:
Test Condition:
Frequency : 1KHz
Level : 1Vrms
DUT : 100nF
Then
=
=
=
1590
9
10100
3
102
1
2
1
π
π
Cxf
Zx
Refer to the accuracy table, get C
Ae
=±0.1%
L Accuracy:
LxfZx =
π
2
L
Ae
= Ae of L
f : Test Frequency (Hz)
Lx : Measured Inductance Value (H)
|Zx| : Measured Impedance Value ()
Accuracy applies when Dx (measured D value) 0.1
When Dx > 0.1, multiply L
Ae
by
2
1 Dx+
Example:
Test Condition:
Frequency : 1KHz
Level : 1Vrms
DUT : 1mH
Then
=
=
=
283.6
3
10
3
102
2
π
π
LxfZx
Refer to the accuracy table, get L
Ae
= ±0.5%