Bk-precision 4078 - Manual Manual do Utilizador Página 69

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 103
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 68
:SOURce <MANual|INTernal|EXTernal|BUS
:TIMer <numeric value>
Note: For model 4078, nothing changes in the commands above to control channel 1. But for channel 2, change
:TRIG to :TRIG2. For example, to change channel 2 mode to gate, send the command: TRIG2:MODE
GATE.
4.13.3.1 Trigger Mode
:TRIGger:MODE <trigger mode>
This command is used to set the trigger mode. It is not a standard SCPI command.
Arguments
Type: Character
Options: CONTinuous
TRIGger
GATE
BURSt
Command Type: Setting or Query
Setting
Syntax: :TRIGger:MODE<ws><option>
Examples: :TRIG:MODE CONT
:TRIG:MODE BURS
Query
Syntax: :TRIGger:MODE?
Response: CONT|TRIG|GATE|BURS
4.13.3.2 Trigger Source
:TRIGger:SOURce <trigger source>
This command is used to select the trigger source, for use in the Trigger, Gate and Burst trigger modes.
Arguments
Type: Character
Options: MANual - Front panel MAN key
BUS - GPIB trigger (GET or *TRG)
INTernal - Internal trigger
EXTernal - External trigger
Command Type: Setting or Query
Setting
Syntax: :TRIGger:SOURce<ws><option>
Examples: :TRIG:SOUR BUS
:TRIG:SOUR INT
Query
Syntax: :TRIGger:SOURce?
Response: MAN|BUS|INT|EXT
4.13.3.3 Burst Count
:TRIGger:BURSt <burst count>
Used to set the number of cycles to be output in the BURST mode. It is not a standard SCPI command.
Vista de página 68