Arguments:
Type: NRf
Units: S (seconds), MS (milliseconds), US (microseconds), NS nanoseconds
Range: 100NS to 99.99S
Rounding: To current resolution
Examples: :TRIG:TIM 10E-6
:TRIG:TIM 500US
QUERY
Syntax: :TRIGger:TIMer?
Examples: :TRIG:TIM?
Response: NR3
4.12.3.5 External Trigger Level
Used to control the trigger level of the external trigger.
COMMAND TYPE: Setting or Query
SETTING
Syntax: :TRIGger:LEVel<ws><trigger level>[units]
Arguments:
Type: NRf
Units: V, mV
Range: 10V to +10V with 10mV resolution; 0V allowed
Rounding: 10mV
Examples: :TRIG:LEV 5.56
QUERY
Syntax: :TRIGger:LEVel?
Examples: :TRIG:LEV?
Response: NR3
4.12.3.6 Trigger Slope
This command is used to set the external trigger slope on which to trigger.
COMMAND TYPE: Setting or Query
SETTING
Syntax: :TRIGger:SLOPe<ws><Options>
Arguments:
Type: Character
Options: POSitive
NEGative
Examples: :TRIG:SLOP POS
:TRIG:SLOP NEG
QUERY
Syntax: :TRIGger:SLOPe?
Examples: :TRIG:SLOP?
Response: POS | NEG